1. GENERAL REQUIREMENTS
|
-
|
|
|
I. Vendors must comply with all the requirements specified in this tender document.
|
Ya Tidak
|
|
|
II. Vendors must bear all costs of supplying the specified quantity of items to UiTM. Such costs shall include the costs of delivery, installation, commissioning, training, troubleshoot and maintenance work (including parts labour and on-site service) within the specified warranty period
|
Ya Tidak
|
|
|
III. Vendors must bear all costs of supplying the specified quantity of items to UiTM. Such costs shall include
the costs of delivery, installation, commissioning, troubleshoot and maintenance work (including parts,
labour and on-site service) within the specified warranty period
The equipment will be distributed in the following Campuses:
1. Electrical Engineering School, College of Engineering, UiTM Shah Alam Selangor x 1 Sets
2. UiTM Penang Branch, Permatang Pauh Campus x 1 Sets
|
Ya Tidak
|
|
|
2. SPECIFICATION
|
-
|
|
|
I. GENERAL SPECIFICATIONS:
- Minimum Six built-in modulation types and 17 popular waveforms to simulate typical applications for testing
- Minimum 7-inch-wide video graphics array (WVGA) color display for a simultaneous parameter set up, signal viewing and editing
- Acquire 16-bit arbitrary waveform capability with memory up to 8 M samples per channel
- Using the USB and LAN IO interface for remote connectivity
- Use BenchVue software to enable PC control
- 20 MHz sine and square waveforms
- SCPI (Standard Commands for Programmable Instruments) compatibility
- Number of channel: 2 Channels (function/arbitrary waveform generator)
- Max Bandwith: 20 MHz
- Max frequency: 20 MHz
- Max Arbitrary waveform length: 1 MSa/S
- Max Sample Rate: 250 MSa/S
- Total Harmonic Distortion: < 0.075% (Typical)
|
Ya Tidak
|
|
|
II. WAVEFORM:
- Standard: Sine, Square, Ramp, Pulse, Triangle, Gaussian Noise, PRBS (Pseudorandom Binary Sequence), DC
- Built-in arbitrary: Cardiac, Exponential Fall, Exponential Rise, Gaussian Pulse, Haversine, Lorentz, D-Lorentz, Negative Ramp, Sinc
- User-defined arbitrary: Up to 8 MSa per channel deep memory; with up to 1Msa per waveform
- Frequency accuracy (standard):
± (1 ppm of setting + 15 pHz), 1 year, 23 ºC ± 5 ºC
± (2 ppm of setting + 15 pHz), 1 year, 0 ºC to 55 ºC
|
Ya Tidak
|
|
|
III. OPERATING MODES AND MODULATION TYPES:
- Operating modes: Continuous, Modulate, Frequency Sweep, Gated Burst
- Modulation types: Amplitude modulation (AM), frequency modulation (FM), phase modulation (PM), frequency shift keying (FSK), binary phase shift keying (BPSK), pulse width modulation (PWM)
|
Ya Tidak
|
|
|
IV. TWO-CHANNEL CHARACTERISTICS
- Operating modes: Independent, coupled parameter(s), combined (Channels 1 and 2); equal (Channel 1 = Channel 2) or differential (Ch 1 = -Ch 2)
- Parameter coupling: None, frequency (ratio or difference) and/or amplitude and DC offset
- Relative phase: 0° to 360°, 0.1° resolution
- Channel-to-channel skew (typical); both channels configured identically: < 0.8 ns
- Crosstalk (typical): < -75 dB
Sync out/trigger out
- Connector : Front BNC, chassis-referenced; functions as an output
- Minimum output high voltage : Minimum 1.3 V
- Maximum output low voltage : Maximum 0.1V
|
Ya Tidak
|
|
|
V. MEMORY
Instrument State
- Store/ recall: User-defined instrument states with user-defined names in the file system
- Power-on state: Default settings or state at power-off, selectable
USB file system
- Front-panel port: USB 2.0 high-speed mass storage class (MSC) device
- Capability: Read or write instrument configuration settings, instrument states, arbitrary waveform
- Speed (nominal): 10 MB/s
|
Ya Tidak
|
|
|
VI. GENERAL CHARACTERISTICS
USB file system
- LXI-C (rev1.5): 10/100Base-T (sockets and VXI-11 protocols); USB 2.0 (USB-TMC488 protocol)
- Web user interface: Remote operation and monitoring
- Programming language: SCPI-1999, IEEE-488.2
- Real-time clock / calendar battery: CR-2032 coin type, replaceable, > 5-year life (typical)
Environmental
- Storage temperature: -40 °C to 70 °C
- Warm-up time: 1 hour
- Operating environment: Indoor use, installation category II for AC input; pollution degree 2
- Operating temperature: 0 °C to 55 °C
- Operating humidity: Up to 80% RH at 40 °C non-condensing
- Altitude: Up to 3,000 meters (9842.5 ft)
Line power
- Line voltage: 100 to 240 V, 50 / 60 Hz; 100 to 120 V, 50 / 60 Hz
- Power consumption: < 45 W
|
Ya Tidak
|
|
|
VII. SUPPLIED ACCESSORIES
-AC power cord
|
Ya Tidak
|
|
|
3. CATALOGUES / BROCHURES
|
-
|
|
|
I. Vendors must provide the catalogues / brochures / technical specification manuals with the tender proposal.
|
Ya Tidak
|
|
|
II. Original copies catalogue of user, brochure, operational manual and clear technical specification of the equipment must be provided
|
Ya Tidak
|
|
|
III. The equipment offered by vendor must be of the latest available model.
|
Ya Tidak
|
|
|
4. TRAINING
|
-
|
|
|
I. Tenderers are required to provide training for the use, maintenance and troubleshoot of the equipment to the Technical Staff & Lecture of Electrical Engineering School, College of Engineering, UiTM Shah Alam and all training-related costs are borne by the tenderers.
|
Ya Tidak
|
|
|
II. A minimum of one (1) day training for a minimum of one (1) participant covering each module. (Vendors must provide details/syllabus of the training module)
|
Ya Tidak
|
|
|
III. Complete training materials and documentation must be provided to all participants
|
Ya Tidak
|
|
|
IV. The training MUST be conducted by trainers who are certified by principal. Please attach the trainer's certification that is certified by principal supplier.
|
Ya Tidak
|
|
|
V. Certificate of attendance must be provided
|
Ya Tidak
|
|
|
VI. Training sessions must be conducted at the following Campuses:
1. School of Electrical Engineering, College of Engineering, UiTM Shah Alam Selangor
2. UiTM Penang Branch, Permatang Pauh Campus
|
Ya Tidak
|
|
|
5. WARRANTY AND SUPPORT
|
-
|
|
|
I. Warranty for the equipment is expected to be at least One (1) year warranty.
|
Ya Tidak
|
|
|
II. Installation, maintanance, testing and commissioning of the entire system should be completed within (12) week from the delivery date.
|
Ya Tidak
|
|
|
III. Vendors must have local service for repair services and provide on call services when required by UiTM
|
Ya Tidak
|
|
|
IV. Vendors must provide details of warranty. INCLUDING ALL parts, labour, basic phone support for hardware issue, repairs onsite) Please attach the Letter of Warranty (LOW) from the principle, which states that UiTM is entitled to 3 years of warranty upon project completion
|
Ya Tidak
|
|
|
V. Vendors be responsible for the installation, commissioning and test-run of the equipment and accompanying software (if applicable) to the satisfaction of UNIVERSITI TEKNOLOGI MARA before final acceptance.
|
Ya Tidak
|
|
|
VI. All cost on delivery. installation, commissioning and test run shall be borne by the vendor.
|
Ya Tidak
|
|
|
VII. The vendors MUST provide a genuine product. The vendors must be supported back-to-back by a principle (manufacturer) to provide maintenance services for equipment supplied. The vendors should attach a Letter of Authorization (LOA) from Distributors/ Principal.
|
Ya Tidak
|
|
|
VIII. Free support through fax, email, telephone must be provided during the warranty period
|
Ya Tidak
|
|
|
6. OTHER REQUIREMENTS
|
-
|
|
|
I. Note that all other related material, components and cabling, which are required to ensure full operation of the above package must be supplied and borne by vendors
|
Ya Tidak
|
|