ADS-SHR-DOC-SPECTECH
  • UiTM/B2/PER/SHR/B/0625/0007
  • CADANGAN MEMBEKAL, MENGHANTAR, MEMASANG, MENGUJILARI DAN MENYELENGGARA PETI SEJUK BEKU -86°C DI ATTA-UR-RAHMAN INSTITUTE FOR NATURAL PRODUCT DISCOVERY (AuRIns), UiTM CAWANGAN SELANGOR KAMPUS PUNCAK ALAM
  • Chiller/Freezer/Lab Refrigerator/Coolroom/Freezer Room
  • AuRIns - Ultra-Low Temperature Freezer (-86°C)
  • 1 Set

Maklumat Spesifikasi Teknikal

  
Keterangan Calit Tawaran/Catatan
1. All items must be delivered and installed at the Bioassay Laboratory, AuRIns, Level 9, Building FF3, UiTM Puncak Alam Campus Ya

Tidak
2. Freezer Specification -
I. Cabinet: Vertical/upright
Ya

Tidak
II. Direct cooling
Ya

Tidak
III. Temperature range: -40°C ~ -86°C
Ya

Tidak
IV. Capacity: 828L and above
Ya

Tidak
V. Exterior Dimention (MUST NOT EXCEED THIS MEASUREMENT)
Height: < 1980mm
Length: < 1200mm
Width: < 1000mm

*This is the maximum size that can fit in the elevator at Building FF3 (lif Bomba)
Ya

Tidak
VI. Controller: Microprocessor
Ya

Tidak
VII. Display: LED
Ya

Tidak
VIII. Alarm (Sound, Light, Remote)
  1. High/Low temperature alarm
  2. Sensor error
  3. Low battery
  4. Door ajar
  5. power failure
  6. Hot condensor
  7. High ambient
Ya

Tidak
IX. Inner doors: 4 (individual and insulated)
Ya

Tidak
X. Shelves: 4 including rack accessory for 1 shelf
Ya

Tidak
XI. Must have USB Port and Interface for data monitoring
Ya

Tidak
XII. Must have door heating system
Ya

Tidak
XIII. Can be locked for security purposes
Ya

Tidak
XIV. Power Supply : approximately at 220 V - 240 V / 50 Hz
Ya

Tidak
XV. Low energy consumption
Ya

Tidak
XVI. Silent type compressor : must be less than 55dB
Ya

Tidak
XVII. Must come with data logger with real time monitoring and can send real time alarm/notification to phone or email
Ya

Tidak
XVIII. Must include Uninterruptible Power Supply (UPS) as battery backup (can supply for at least 30 minutes)
Ya

Tidak
XIX. Temperature accuracy: 0.1°C
Ya

Tidak
XX. Brand: Haier or equivalent
Ya

Tidak
3. Warranty : 5 years parts and labour, including one preventive maintenance service every year of the warranty. Ya

Tidak
4. Training and commissioning must be done during delivery after installation Ya

Tidak
5. Installation of freezer must be done along with data logger and UPS Ya

Tidak
6. Simple Standard Operational Procedure (SOP) must be included. Ya

Tidak