ADS-SHR-DOC-SPECTECH
  • UiTM/B1/PER/SHR/B/0625/0038
  • Cadangan Membekal, Menghantar, Memasang, Mengujilari Dan Menyelenggara Peralatan Pengajaran dan Pembelajaran Untuk Makmal Elektronik (Semikonduktor 13), Fakulti Kejuruteraan Elektrik, UiTM Shah Alam
  • Electrical/Electricity/Electronic Wares - Multimeter/Multitester/Ammeter/Voltmeter
  • PKE 2025 (V2) 6½-Digit Dual-Display Digital Multimeter
  • 15 Set

Maklumat Spesifikasi Teknikal

  
Keterangan Calit Tawaran/Catatan
1. General -
I. Measures 11 input signals:
  • DC voltage,
  • DC current,
  • True RMS AC voltage,
  • AC Current,
  • Two- and four-wire
  • Resistance,
  • Frequency,
  • Continuity,
  • Diode test,
  • Temperature,
  • Capacitance
Ya

Tidak
II.
  • Signature 7-inch dual-measurement color display
  • Fast reading rate of up to 1000 readings/s for speed-critical measurements
  • 10,000 points logging memory for recording more data and perform analysis or better
  • Standard USB and LAN for flexible PC connectivity
  • USB flash drive support to copy / load configuration for repeated test setup
  • Includes digital multimeter (DMM) software for remote control and data logging
Ya

Tidak
III.
  • Combination of hard and soft keys for more intuitive front-panel operation
  • Standard connectivity to the PC
  • Signature 7-inch dual-measurement color display
  • USB flash drive adds convenience
Ya

Tidak
IV.
  • Simplify Data Gathering and Analysis with digital multimeter (DMM) Software
  • Record measurements and export results in a few clicks
  • Remotely control your DMM via LAN
  • Power supply: 100 V / 120 V (127 V) / 220 V (230 V) / 240 V ± 10%, AC line frequency 45 Hz – 66 Hz, automatically sensed at power on
  • Shock and vibration: Tested to IEC/EN 60086-2
Ya

Tidak
2. DC specifications Accuracy ± (% of reading + % of range): -
I. DC voltage
  • Range: 100.000 mV (min)
  • Input impedance: 10 M ohm or > 10 G ohm
  • Temperature coefficient: 0.0005 + 0.0005
Ya

Tidak
II.
  • Range: 1000.00 V (max)
  • Input impedance: 10 M ohm
  • Temperature coefficient : 0.0005 + 0.0001
Ya

Tidak
3. Resistance -
I.
  • Range: 100.000 ohm (min)
  • Test current or burden voltage: 1 mA
  • Temperature coefficient : 0.0006 + 0.0005
Ya

Tidak
II.
  • Range: 100.000 M ohm (max)
  • Test current or burden voltage: 500 nA || 10 M?
  • Temperature coefficient : 0.1500 + 0.0002
Ya

Tidak
4. DC current -
I.
  • Range: 100.0000 µA (min)
  • Test current or burden voltage: < 0.011 V
  • Temperature coefficient : 0.0020 + 0.0030
Ya

Tidak
II.
  • Range: 10.0000 A (max)
  • Test current or burden voltage: < 0.5 V
  • Temperature coefficient : 0.0050 + 0.0010
Ya

Tidak
5. Continuity -
I.
  • Range: 1000 ohm
  • Test current or burden voltage: 0.5 mA
  • Temperature coefficient : 0.0010 + 0.0020
Ya

Tidak
6. Diode test -
I.
  • Range: 5.0000 V
  • Test current or burden voltage: 0.5 mA
  • Temperature coefficient : 0.0010 + 0.0020
Ya

Tidak
7. True RMS -
I. True RMS AC voltage
100 mV, 1 V, 10 V, 100 V, and 750 V ranges
Ya

Tidak
A )
  • Frequecy : 3 Hz to 5 Hz
  • Temperature coefficient/°C : 0.100 + 0.003
Ya

Tidak
B )
  • Frequency: 5 Hz to 10 Hz
  • Temperature coefficient / ºC : 0.035 + 0.003
Ya

Tidak
C )
  • Frequency: 10 Hz to 20 kHz
  • Temperature coefficient / ºC : 0.005 + 0.003
Ya

Tidak
D )
  • Frequency: 20 kHz to 50 kHz
  • Temperature coefficient / ºC : 0.011 + 0.005
Ya

Tidak
E )
  • Frequency: 50 kHz to 100 kHz
  • Temperature coefficient / ºC : 0.060 + 0.008
Ya

Tidak
F )
  • Frequency: 100 kHz to 300 kHz
  • Temperature coefficient / ºC : 0.200 + 0.020
Ya

Tidak
II. True RMS AC current
100 µA, 1 mA, 10 mA, and 100 mA ranges
Burden voltage : < 0.011, < 0.11, < 0.05, < 0.5 V
Ya

Tidak
A )
  • Frequency: 3 Hz to 5 kHz
  • Temperature coefficient / ºC : 0.015 + 0.006

Ya

Tidak
B )
  • Frequency: 5 kHz to 10 kHz (typ)
  • Temperature coefficient / ºC : 0.030 + 0.006
Ya

Tidak
8. Frequency accuracy ± (% of reading): -
I. Frequency
-
A )
  • Range: 100.000 mV to 750.00 V (min)
  • Frequency: 20 Hz – 300 kHz (min)
  • Temperature coefficient / ºC, 0 ºC – 18 ºC, 28 ºC – 55 ºC: 0.005
Ya

Tidak
B )
  • Range: 10.0000 mA to 3.0000 A (max)
  • Frequency: 20 Hz – 10 kHz (max)
  • Temperature coefficient / ºC, 0 ºC – 18 ºC, 28 ºC – 55 ºC: 0.005
Ya

Tidak
9. Temperature and capacitance specifications Accuracy ± (% of reading + % of range) -
I. Temperature

-
A )
  • PT100 (DIN/IEC 751) : Probe accuracy + 0.05 °C
  • Thermistor : Probe accuracy + 0.1 °C
Ya

Tidak
II. Capacitance
-
A )
  • Range: 1.000 nF (min)
  • Temperature coefficient / ºC: 0.05 + 0.05
Ya

Tidak
B )
  • Range:100 µF (min)
  • Temperature coefficient / ºC: 0.05 + 0.01
Ya

Tidak
10. Supplied Accessories -
I.
  • Power cord Per Unit
  • Test leads Per Unit
Ya

Tidak
11. Manual and Documentation -
I. Original copies of user and operational manual of the equipment must be provided
Ya

Tidak
12. Installation , Testing and Commissioning -
I.
  • Tenderers is required to install and commissioning the proposed equipment in the specified laboratory location in Faculty of Electrical Engineering, UiTM Shah Alam by their qualified trainers.
  • Installation, testing and commissioning of the entire system should be completed within one month from the delivery date.
Ya

Tidak
13. Training -
I. Tenderers are required to provide training for the use, maintenance and troubleshoot of the equipment to the Technical Staff & Lecturer of Faculty of Electrical Engineering, UiTM Shah Alam and all training-related costs are borne by the tenderers
Ya

Tidak
14. After Sales Service -
I.
  • Tenderers must have local service for repair services and provide on call services when required by UiTM.
  • Tenderers must provide details of the warranty ( warranty 3 years)
Ya

Tidak