1. General
|
-
|
|
|
I.
- Bandwidth: 200 MHz
- Channels: 2-channel
- Sample Rate: 2 GS/s sample rate on all channels or better
- Record Length: 2M point record length on all channels or better
|
Ya Tidak
|
|
|
II.
- 7-inch TFT LCD WVGA
- Automatic measurements 14 amplitude, 14 timing, and 4 pulse count measurements
|
Ya Tidak
|
|
|
III.
- USB 2.0 host port on the front panel for quick and easy data storage.
- USB 2.0 device port on rear panel to connect to a PC and remotely control the instrument
- Provides helpful on-screen tips for users
- Built-in oscilloscope handbook provides operating instructions and oscilloscope fundamentals
- Integrated courseware feature provides lab exercise guidance on the display
- Autoset, Cursors, and Automated measurements can be disabled to help educators to teach basic concepts to students
|
Ya Tidak
|
|
|
2. Vertical system
|
-
|
|
|
I.
- Vertical resolution: 8 bits
- Input sensitivity range: 500 μV/div to 10 V/div
- DC gain accuracy: +3% full scale (≥ 10 mV/div) +4% full scale (< 10 mV/div)
- Maximum Input Voltage: 150 Vrms, 200 Vpk
- Offset range: 500 uV/div to 200 mV/div: +2 V , > 200 mV/div to 10 V/div: +100 V
- Bandwidth limit: 20 MHz (Typ)
- Input coupling: DC, AC (10 Hz cutoff frequency)
- Input impedance: 1 MΩ ± 2%, 16 pF ±3 pF
- Vertical zoom: Vertically expand or compress a live or stopped waveform
|
Ya Tidak
|
|
|
3. Horizontal system
|
-
|
|
|
I.
- Timebase accuracy: 50 ppm ± 5 ppm per year (aging)
- Timebase range: 2 ns/div to 50 s/div
- Horizontal zoom: Horizontally expand or compress a live or stopped waveform
- Deskew range: ± 100 nsec
|
Ya Tidak
|
|
|
4. Trigger system
|
-
|
|
|
I.
- External trigger input: Included
- Trigger modes: Auto, Normal, Single, Force
- Trigger Types (Edge): Trigger on a rising, falling, alternating or either edge of any source
- Trigger Types (Pulse Width): Trigger on a pulse of a selected channel with a time duration that is 'less than a value,’ 'greater than a value’ or 'inside a time range’ Range minimum: 10 ns, 10 s max
- Trigger source: CH1, CH2, AUX IN, AC Line
- Trigger Coupling: DC, Noise Reject, High Frequency Reject, Low Frequency Reject
- Trigger signal frequency readout: Provides a frequency readout of the trigger source up to instrument bandwidth
|
Ya Tidak
|
|
|
5. Waveform measurements
|
-
|
|
|
I.
- Cursors: Time, Amplitude, Screen
- Automated measurements: 32, of which up to 4 can be displayed on-screen at any one time. Measurements include: Period, Frequency, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase,
- Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle
- RMS, Positive Pulse Count, Negative Pulse Count, Rising Edge Count, Falling Edge Count, Area, Cycle Area, Delay FR,
- Delay FF, Delay FR, and Delay RR.
- Gating: Isolate the specific occurrence within an acquisition to take measurements on, using either the screen, between waveform cursors or full record length
|
Ya Tidak
|
|
|
6. Waveform math
|
-
|
|
|
I.
- Arithmetic: Add, subtract, multiply, divide, FFT (magnitude), FFT (phase), low-pass filter
- FFT: Window types: Hanning, Flat top, Rectangular, Blackman-Harris Vertical scaling: dB (logarithmic) or RMS (linear) Horizontal scaling: User-defined span and center frequency settings, or Auto Setup
|
Ya Tidak
|
|
|
7. Display system
|
-
|
|
|
I.
- Display Type: 7-inch TFT Color Display or better
- Display Resolution: 800 horizontal by 480 vertical displayed pixels (WVGA)
- Waveform styles: Vectors, Variable persistence, and Infinite persistence.
- Format: YT and XY
|
Ya Tidak
|
|
|
8. Supplied Accessories
|
-
|
|
|
I.
- Probe set Per Unit
- AC Plug Per Unit
|
Ya Tidak
|
|
|
9. Manual and Documentation
|
-
|
|
|
I. Original copies of user and operational manual of the equipment must be provided
|
Ya Tidak
|
|
|
10. Installation , Testing and Commissioning
|
-
|
|
|
I.
- Tenderers is required to install and commissioning the proposed equipment in the specified laboratory location in Faculty of Electrical Engineering, UiTM Shah Alam by their qualified trainers.
- Installation, testing and commissioning of the entire system should be completed within one month from the delivery date.
|
Ya Tidak
|
|
|
11. Training
|
-
|
|
|
I. Tenderers are required to provide training for the use, maintenance and troubleshoot of the equipment to the Technical Staff & Lecturer of Faculty of Electrical Engineering, UiTM Shah Alam and all training-related costs are borne by the tenderers
|
Ya Tidak
|
|
|
12. After Sales Service
|
-
|
|
|
I.
- Tenderers must have local service for repair services and provide on call services when required by UiTM.
- Tenderers must provide details of the warranty. (warranty 3 years)
|
Ya Tidak
|
|