  • CADANGAN PERKHIDMATAN MENYELENGGARA LIQUID CHROMATOGRAPHY QTRAP MASS SPECTROMETRY AB SCIEX 6500 DAN MEMBEKAL, MENGHANTAR BAHAN PAKAI HABIS DAN NITROGEN GENERATOR KE FAKULTI SAINS GUNAAN, UiTM SHAH ALAM, SELANGOR (GERAN IMAP (007/2024))
  • Liquid Chromatography-Mass Spectrometry (LC-MS)
  • Penyelenggaraan Peralatan Makmal Liquid Chromatography-Mass Spectrometry Quadrapole TRAP (LC/MS Q-Trap) AB Sciex 6500
  • 1 Set

Maklumat Spesifikasi Teknikal

  
Keterangan Calit Tawaran/Catatan
1. Repair and maintenance service works for AB Sciex QTrap LCMSMS system Ya

Tidak
2. To carry out Pre-PM pressure test and system test Ya

Tidak
3. To replace the roughing pump oil and exhaust filter Ya

Tidak
4. To replace the four air filters in the base of the mass spectrometer chassis. Ya

Tidak
5. To clean the turbo pump filter screen and inspect the mass spectrometer, components, and cabling, as required Ya

Tidak
6. To clean curtain plate, orifice plate, IonDriveTM QJet ion guide, IQ0 lens and Q0 region Ya

Tidak
7. To replace the TurboIonSpray® or APCI electrodeg) To verify that the temperature (TEM) reaches a set point (such as 200°C). Ya

Tidak
8. To carry out Post-PM pressure test and system test such as Q1 and Q3 High Mass Positive PPG, Q1 and Q3 HighMass Negative PPG, Positive Enhanced Resolution, Negative Enhanced Resolution, Low Mass Q1 and Q3 Positive PPG,Low Mass Q1 and Q3 Negative PPG and Enhanced Product Ion Mode test. Ya

Tidak
9. Record verification data in the PM document and then attach all performance dataj) For Eksigent110XL UHPLC, replace the solvent filters in the solvent bottles. Ya

Tidak
10. Training after maintenance Ya

Tidak
11. To replace the high pressure seals on all four pump heads, using the tool in the seal kit Ya

Tidak
12. To replace the plunger seals on all four pump heads, using the tool in the seal kit Ya

Tidak
13. To replace the outlet filter on each pump n) To replace the injection valve rotor seal. Ya

Tidak
14. To replace the syringe valve, buffer tubing, air needle and sample needle Ya

Tidak
15. To confirm the sample needle position at the sample vial. Ya

Tidak
16. Monitor the temperature on the column oven display to confirm that the set point is reached. Ya

Tidak
17. To perform the pressure test and record results in the Planned Maintenance Data Log Ya

Tidak
18. To perform the HPLC MS system test and record results in the Planned Maintenance Data Log Ya

Tidak
19. For nitrogen generator, replace and refit all compressor with service kits. Replace all filters and check for any leaking Ya

Tidak
20. Review the routine maintenance schedule and procedures with the customer Ya

Tidak
21. Warranty cover 12 months and should be cover for training after services Ya

Tidak